How to Calibrate an RF Low-Noise Amplifier Test System
1. Preparing for Calibration
Equipment Status Check
-
Verify all test system equipment is functional
-
Confirm signal generator, power meter, and noise figure analyzer are operational
-
Ensure equipment has passed regular measurements
Environmental Setup
-
Place system in constant temperature/humidity environment
-
Avoid strong electromagnetic interference
-
Prevent environmental factors affecting calibration accuracy
Standard Components Preparation
-
Prepare standard attenuators and noise sources
-
Verify parameters are accurate and within validity period
-
Ensure reliable reference for subsequent calibrations
2. Signal Chain Calibration
Signal Amplitude Calibration
-
Directly connect signal generator to power meter
-
Output standard RF signals of different frequency bands
-
Record power meter measurements and compare differences
-
Adjust equipment parameters to align with standard values
Passive Components Measurement
-
Connect RF cables and connectors to test system
-
Measure signal attenuation through components
-
Incorporate attenuation values into test data corrections
-
Ensure accuracy of signal chain transmission characteristics
Noise Figure Analyzer Calibration
-
Use standard noise source to calibrate noise figure analyzer
-
Set up analyzer according to standard noise source parameters
-
Measure output noise power of standard noise source
-
Compare measured values with theoretical values
-
Adjust analyzer calibration factor to minimize deviation
RF Low Noise Amplifier Testing
-
Connect RF low-noise amplifier to test link
-
Use calibrated noise figure analyzer and standard attenuator
-
Correct for test link's own noise contribution
-
Ensure accuracy of amplifier's noise figure measurement
4. Calibration Verification & Documentation
Verification Process
-
Select standard RF low-noise amplifier with known performance
-
Connect to calibrated test system for testing
-
Compare test results with standard parameters
-
Validate calibration if deviation is within specified range
Documentation Requirements
-
Record device parameters and calibration steps
-
Document measurement data and deviations
-
Create complete calibration record for traceability
-
Ensure test system maintains stable calibration state